Size: |
212 mm (W), 168 mm (D), 313 mm (H)
(8.3" (W), 6.6" (D), 12.3" (H)
|
Unit weight: |
4.66 kg (10.25 Ibs) (without carry bag and cart) |
Power requirements: |
AC adaptor: 100-240V AC(+/- 10%) 50-60 Hz in
24V DC, 6.25A out
DC adaptor: 11.5 - 16V DC in, 19V, 7.9A out
(Important: see accessories list for model and part number of AC power supply.)
|
Purity: |
87% - 96% at all flow rates, over operating conditions |
Setting: |
User adjustable in 0.5 increments from 1.0 to 6.0 in pulse mode and from 0.5 to 2.0 in continuous mode. |
Inspiratory trigger sensitivity: |
-0.12cm/H20 |
Setting indicator: |
LCD display |
Maximum oxygen discharge pressure: |
20.5 psi |
Humidity range: |
5% to 93% ± 2% non-condensing |
Operating altitude: |
0’ to 13,000’ (0m to 4,000m) relative to sea level, 1,060 down to 575 mbar |
Sound level: |
Sound Pressure level of 38 dB(A) at setting 2 in pulse mode, tested according to Prufmethode 14-1 03/2007 MHS-Hi
Sound Pressure level of 46dBA/Sound Power level of 54 dBA at setting 6 in pulse mode
Sound Pressure level of 52 dBA/Sound Power level of 60 dBA at setting 2 continuous
|
Type of protection (electrical): |
Class II |
Degree of protection (electrical): |
Type BF
|
Degree of protection (water): |
IP22 in carry bag (protection against small objects and tilted drip-ping water)
IP20 out of carry bag (protection against small objects and no protection against water entering the concentrator)
|
Degree of safety (flammable anaesthetic mixture): |
Not suitable for use in the presence of a flammable anaesthetic mixture
|
Operating temperature: |
Continuous operating at temperatures between 5°C (41°F) and 40°C (104°F).
|
Storage temperature: |
Between -20°C (-4°F) and 60°C (140°F).
|
Alarm sound pressure level: |
Less than 10 second after detection (low oxygen alarms if oxygen is less than 82% volume fraction at specified environmental conditions)
|
Oxygen concentrator status indicator: |
High priority alarm that indicates when oxygen concentration drops below 82%.
|